HV-25 and HV-50 Specifications


MODEL
HV-25
HV-50
TEMPERATURE
Operating Range
105ºC to 126ºC
105ºC to 135ºC
Display Range
128ºC
137ºC
Warming Range
45ºC to 60ºC
Agar Dissolution
Heating at 60ºC to 100ºC for up to 1 hour (temperature and time programmable), followed by 45ºC to 60ºC
(programmable for up to 20 hours to prevent coagulation)
GAUGE PRESSURE
Operating Range
3 to 22 psi gauge (0.02 to 148MPa)
3 to 30 psi (0.02 to 0.206MPa)
Maximum Allowable
27 psi (0.186MPa)
38 psi (0.26MPa)
TIMER
Sterilization
1 to 250 minutes
Delayed Auto-start
Program autoclave to start 1 minute to 7 days later
DIMENSIONS
Overall (W x D x H)
19.0" x 18.0" x 37.4"
(48 cm x 46 cm x 95 cm)
20" x 21" x 41"
(51 cm x 53 cm x 104 cm)
Chamber
(Diameter x Height)
25 liters
9.4" x 21.6" (24 cm x 55 cm)
50 liters
11.8" x 28.0" (30 cm x 71 cm)
SAFETY DEVICES
Dual-sensing lid interlock, over-pressure power cutoff, over-temperature power cutoff, low-water power cutoff, mechanical pressure relief valve, microprocessor-controlled pressure relief valve, lid-closure detector, over-current detector and circuit breaker
PROGRAMS
1. For agar
Heating -> Sterilization -> Exhaust (from no exhaust to slow, programmable) -> Warming -> Completion
2. For liquids
Heating -> Sterilization -> Exhaust (from no exhaust to slow, programmable) -> Completion
3. For solids
Heating -> Sterilization -> Exhaust (fast, default) -> Completion
4. For melting agar or for faster start-up
Rapid melting at 60°C to 100°C (temperature and time programmable) -> Maintain at 45°C to 60°C (temperature programmable) -> Completion
ACCESSORIES
Stainless steel
Mesh wire baskets, buckets (pails), and round test tube racks
Factory or field installed
Floating (load) sensor, digital printer, and data logger
POWER
Source
120V, 50/60 Hz (220V/240V available)
Consumption
1500 watts (1500VA)
2000 watts (2000VA)
SHIPPING WEIGHT
160 lb (73 kg)
180 lb (82 kg)

WARRANTY: Autoclaves are protected by a 12-month warranty.