HVA-85 and HVA-110 Specifications
MODEL | HVA-85 |
HVA-110 |
TEMPERATURE | ||
Operating Range | 105ºC to 135ºC |
|
Display Range | 137ºC |
|
Agar Warming Range | After sterilization, agar can be maintained from 45ºC to 60ºC (programmable) to prevent coagulation for up to 20 hours. |
|
GAUGE PRESSURE | ||
Maximum Allowable | 38 psi (0.26MPa) |
|
Operating Range | 3 to 30 psi (0.021 to 0.206MPa) |
|
TIMER | ||
Sterilization | 1 to 250 minutes |
|
DIMENSIONS | ||
Overall (W x D x H) | 26.0"" x 25.6" x 39.4" (66 cm x 65 cm x 100 cm) |
26.0" x 25.6" x 46.4" in (66 cm x 65 cm x 118cm) |
Chamber (Diameter x Height) |
85 liters 16.5" x 24.2" (42.0 cm x 61.5 cm) |
110 liters 16.5" x 31.3" (42.0 cm x 79.5 cm) |
SAFETY DEVICES | Dual-sensing lid interlock, over-pressure power cutoff, over-temperature power cutoff, low-water power cutoff, mechanical pressure relief valve, microprocessor-controlled pressure relief valve, lid-closure detector, over-current detector and circuit breaker |
|
PROGRAMS | ||
1. For agar | Heating -> Sterilization -> Auto steam exhaust or ambient cooling (programmable) -> Warming -> Completion |
|
2. For liquids | Heating -> Sterilization -> Auto steam exhaust or ambient cooling (programmable) -> Completion |
|
3. For solids | Heating -> Sterilization -> Auto exhaust (fast) -> Completion |
|
ACCESSORIES | ||
Stainless steel | Mesh wire baskets, buckets (pails), and round test tube racks |
|
Factory or field installed | Data logger |
|
POWER | ||
Source | 208V / 220V / 240V, 50 / 60 Hz, single phase |
|
Consumption | 3000 watts (3000VA) |
4000 watts (4000VA) |
SHIPPING WEIGHT | 282 lb (128 kg) |
308 lb (140 kg) |
WARRANTY: Autoclaves are protected by a 12-month warranty.