Autoclaves

HVE-50 Specifications


TEMPERATURE
Operating Range
105ºC to 135ºC
Display Range
5ºC to 137ºC
Warming Range
45ºC to 60ºC
GAUGE PRESSURE
Operating Range (gauge)
3 to 30 psi (0.021 to 0.206MPa)
Maximum Allowable
38 psi ( 0.26MPa)
TIMER
Sterilization
1 to 250 minutes
DIMENSIONS
Overall (W x D x H)
20.1" x 20.8" x 40.9" (51 cm x 53 cm x 104 cm)
Chamber
(Diameter x Height)
50 liters
11.8" x 28" (30 cm x 71 cm)
SAFETY DEVICES
Dual-sensing lid interlock, over-pressure power cutoff, over-temperature power cutoff, low-water power cutoff, mechanical pressure relief valve, lid-closure detector, over-current detector and circuit breaker
PROGRAMS
1. For agar/liquids. Also for melting agar, or for faster start-up
Heating -> Sterilization -> Auto steam exhaust or ambient cooling (programmable) -> Warming -> Completion
2. For liquids/solids
Heating -> Sterilization -> Auto steam exhaust or ambient cooling (programmable) -> Completion
3. For solids
Heating -> Sterilization -> Auto exhaust (fast) -> Completion
ACCESSORIES
Stainless steel
Mesh wire baskets, buckets (pails), and round test tube racks
Factory or field installed
Data logger
POWER
Source
120V, 50/60 Hz, single phase (220V available on request)
Consumption
2000 watts (2000 VA)
SHIPPING WEIGHT
180 lb (82 kg)

WARRANTY: Autoclaves are protected by a 12-month warranty.