Autoclaves and Sterilizers

HVP-50 Specifications


TEMPERATURE
Operating Range
105ºC to 135ºC
Display Range
5ºC to 140ºC
GAUGE PRESSURE
Operating Range (gauge)
3 to 28.4 psi (0.2 to 2.0 kg/cm² )
TIMER
Sterilization
1 to 250 minutes
Drying
1 to 90 minutes
DIMENSIONS
Overall (W x D x H)
20.1" x 24.8" x 40.9" (51 x 63 x 104) cm
Chamber
(Diameter x Height)
50 liter
11.8" x 28" (30 cm x 71 cm)
SAFETY DEVICES
Temperature- and pressure-sensitive lid interlock, over-pressure power cutoff, over-temperature power cutoff, low-water power cutoff, mechanical pressure relief valve, lid-closure detector, over-current detector and circuit breaker with ground fault circuit interrupter.
PROGRAMS
1. For solids with drying
Heating and air removal from chamber by vacuum pump (pre-vac) -> Sterilization -> Water and steam exhaust (fast) -> Drying by heat and vacuum -> Completion
2. For wastes/solids without drying
Heating -> Pre-vac -> Sterilization -> Steam exhaust (fast) -> Completion
3. For liquids/agar
Heating -> Sterilization -> Auto steam exhaust (programmable rates, or ambient cooling w/o steam exhaust) -> Completion
4. Drying without sterilization
Drying with heat and vacuum (time programmable) -> Completion(for additional drying after Program #1, if desired)
ACCESSORIES
Stainless steel
Mesh wire baskets, buckets (pails), and round test tube racks
Factory or field installed
Data logger
POWER
Source
120V, 50/60 Hz, single phase (220V/240V available on request)
Consumption
2000 watts (2000 VA)
SHIPPING WEIGHT
291 lb (133 kg)

WARRANTY: Autoclaves are protected by a 12-month warranty.